Medical device recalls Moderate risk

Microgenics Corporation recalls Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, O…

Recall date
October 1, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0415-2026
FDA classification
Class II
Brand / firm
Microgenics Corporation
Sold / distributed
Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.

Why it was recalled

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control