Microgenics Corporation recalls Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, O…
- Recall date
- October 1, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0415-2026
- FDA classification
- Class II
- Brand / firm
- Microgenics Corporation
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.
Why it was recalled
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control