MicroPort Orthopedics Inc. recalls Cobalt Chromium Heads Transcend: 26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 260000…
- Recall date
- June 21, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0689-2017
- FDA classification
- Class III
- Brand / firm
- MicroPort Orthopedics Inc.
- Sold / distributed
- Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia,…
Why it was recalled
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Class III means the product violates regulations but is not likely to cause harm.
What was recalled
Cobalt Chromium Heads Transcend: 26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 26000025, 26000026, 26000027, 26000028 Conserve: 38013800, 38013804, 38013835, 38014000, 38014004, 38014035, 38014200, 38014204, 38014235, 38014400, 38014404, 38014435, 38014600, 38014604, 38014635, 38014800, 38014804, 38014835, 38015000, 38015004, 38015035, 38015200, 38015204, 38015235, 38015400, 38015404, 38015435, 38015600, 38015604, 38015635, 38AC3600, 38AC3800, 38AC4000, 38AC4200, 38AC4400, 38AC4600, 38AC4800, 38AC5000, 38AC5200, 38AC5400, 38AC5600, 38AM3600, 38AM3604, 38AM3635, 38AM3800, 38AM3804, 38AM3835, 38AM4000, 38AM4004, 38AM4035, 38AM4200, 38AM4204, 38AM4235, 38AM4400, 38AM4404, 38AM4435, 38AM4600, 38AM4604, 38AM4635, 38AM4800, 38AM4804, 38AM4835, 38AM5000, 38AM5004, 38AM5035, 38AM5200, 38AM5204, 38AM5235, 38AM5400, 38AM5404, 38AM5435, 38AM5600, 38AM5604, 38AM5635, PL38HM4400, PL38HM4404, PL38HM4435, PL38HM4600, PL38HM4604, PL38HM4635, PL38HM4800, PL38HM4804, PL38HM4835, PL38HM5000, PL38HM5004, PL38HM5035, PL38HM5200, PL38HM5204, PL38HM5235