Medical device recalls Moderate risk

Midmark Corp dba Progeny Inc recalls Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray…

Recall date
September 4, 2014
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0651-2016
FDA classification
Class II
Brand / firm
Midmark Corp dba Progeny Inc
Sold / distributed
Worldwide Distribution - US including AL, AR, AZ, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI; and Internationally to Australia, Belgium, Canada, Czech Republic, France, Hungary, Lithuania, New Zealand, Poland, Russia, S…

Why it was recalled

Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu

Get recall alerts

Free email alert whenever Midmark Corp dba Progeny Inc has a new recall — straight from official government data. Unsubscribe anytime.