Medical device recalls Moderate risk

MIM Software Inc recalls MIM software; System, Image Processing, Radiological

Recall date
February 19, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1267-2025
FDA classification
Class II
Brand / firm
MIM Software Inc
Sold / distributed
Worldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.

Why it was recalled

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MIM software; System, Image Processing, Radiological