MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK recalls VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator Sys…
- Recall date
- March 7, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1308-2016
- FDA classification
- Class II
- Brand / firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
- Sold / distributed
- Nationwide Distribution including New York, Texas, Florida, and Ohio
Why it was recalled
In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Get recall alerts
Free email alert whenever MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK