Mobius Medical Systems, LP recalls Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation th…
- Recall date
- February 23, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2100-2017
- FDA classification
- Class II
- Brand / firm
- Mobius Medical Systems, LP
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Sweden, Canada, Netherlands, Denmark, Spain, Switzerland, Australia, Austria, Israel, United Kingdom, France
Why it was recalled
Mobius3D version 1.5.0 contained a defect in software code which affects users who perform beam customization, and may lead to a discrepancy in dose calculation between this version of Mobius3D (1.5.0) and the prior version (1.4.2), where none should have been expected.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Get recall alerts
Free email alert whenever Mobius Medical Systems, LP has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Mobius Medical Systems, LP