Medical device recalls Moderate risk

Natus Medical Incorporated recalls Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and…

Recall date
April 22, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1878-2016
FDA classification
Class II
Brand / firm
Natus Medical Incorporated
Sold / distributed
distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.

Why it was recalled

The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.

Get recall alerts

Free email alert whenever Natus Medical Incorporated has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Natus Medical Incorporated