Natus Neurology DBA Excel Tech., Ltd. (XLTEK) recalls natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph
- Recall date
- March 5, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1630-2020
- FDA classification
- Class II
- Brand / firm
- Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
- Sold / distributed
- Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom
Why it was recalled
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.