Medical device recalls

NeuMoDx Molecular Inc recalls NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and o…

Recall date
May 15, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2093-2023
FDA classification
Class III
Brand / firm
NeuMoDx Molecular Inc
Sold / distributed
US Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.

Why it was recalled

Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs in transport medium and bronchoalveolar lavage (BAL) specimens from individuals suspected of COVID-19 by their healthcare provider (HCP). REF Number: 300800

Get recall alerts

Free email alert whenever NeuMoDx Molecular Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: NeuMoDx Molecular Inc