NeuroLogica Corporation recalls Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
- Recall date
- September 14, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0263-2016
- FDA classification
- Class II
- Brand / firm
- NeuroLogica Corporation
- Sold / distributed
- Distributed to the states of NC, OH, MA, and MO.
Why it was recalled
When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
Get recall alerts
Free email alert whenever NeuroLogica Corporation has a new recall — straight from official government data. Unsubscribe anytime.