Medical device recalls Moderate risk

NeuroLogica Corporation recalls Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Recall date
September 14, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0263-2016
FDA classification
Class II
Brand / firm
NeuroLogica Corporation
Sold / distributed
Distributed to the states of NC, OH, MA, and MO.

Why it was recalled

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Get recall alerts

Free email alert whenever NeuroLogica Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: NeuroLogica Corporation