Neusoft Medical System Co recalls Neusoft NeuViz 64 Multi-slice CT Scanner System
- Recall date
- March 8, 2014
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1693-2015
- FDA classification
- Class II
- Brand / firm
- Neusoft Medical System Co
- Sold / distributed
- Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and T…
Why it was recalled
The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Neusoft NeuViz 64 Multi-slice CT Scanner System