Neusoft Medical Systems Co., Ltd. recalls NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole…

Recall date
October 22, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1281-2019
FDA classification
Class II
Brand / firm
Neusoft Medical Systems Co., Ltd.
Sold / distributed
US Nationwide Distribution.

Why it was recalled

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Get recall alerts

Free email alert whenever Neusoft Medical Systems Co., Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Neusoft Medical Systems Co., Ltd.