Neusoft Medical Systems Co., Ltd. recalls NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body co…
- Recall date
- October 22, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1279-2019
- FDA classification
- Class II
- Brand / firm
- Neusoft Medical Systems Co., Ltd.
- Sold / distributed
- US Nationwide Distribution.
Why it was recalled
Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
Get recall alerts
Free email alert whenever Neusoft Medical Systems Co., Ltd. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Neusoft Medical Systems Co., Ltd.