Medical device recalls Moderate risk

New Medical Technologies Gmbh recalls Integra Miltex Cryosolutions Cartridges 4 Pack (23.5g N2O each), Model: C-CA-23, Item: 33517;

Recall date
December 19, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1104-2025
FDA classification
Class II
Brand / firm
New Medical Technologies Gmbh
Sold / distributed
Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA and the country of Canada.

Why it was recalled

During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Integra Miltex Cryosolutions Cartridges 4 Pack (23.5g N2O each), Model: C-CA-23, Item: 33517;

Get recall alerts

Free email alert whenever New Medical Technologies Gmbh has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: New Medical Technologies Gmbh