Medical device recalls Moderate risk

NIPRO Technical Services, Inc. recalls Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity…

Recall date
December 12, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1185-2026
FDA classification
Class II
Brand / firm
NIPRO Technical Services, Inc.
Sold / distributed
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.

Why it was recalled

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters