Medical device recalls Moderate risk

Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy…

Recall date
October 25, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0679-2025
FDA classification
Class II
Brand / firm
Noah Medical
Sold / distributed
US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

Why it was recalled

Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

Get recall alerts

Free email alert whenever Noah Medical has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Noah Medical