Medical device recalls High risk

Nova Biomedical Corporation recalls StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-pat…

Recall date
November 20, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0681-2025
FDA classification
Class I
Brand / firm
Nova Biomedical Corporation
Sold / distributed
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Why it was recalled

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

Get recall alerts

Free email alert whenever Nova Biomedical Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Nova Biomedical Corporation