Medical device recalls High risk

Nova Biomedical Corporation recalls StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone mete…

Recall date
November 20, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0682-2025
FDA classification
Class I
Brand / firm
Nova Biomedical Corporation
Sold / distributed
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Why it was recalled

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing

Get recall alerts

Free email alert whenever Nova Biomedical Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Nova Biomedical Corporation