NRT X-RAY A/S recalls Intelli-C, Right side suspended Tabletop, REF: 03400000
- Recall date
- April 27, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1668-2023
- FDA classification
- Class II
- Brand / firm
- NRT X-RAY A/S
- Sold / distributed
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Why it was recalled
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Intelli-C, Right side suspended Tabletop, REF: 03400000
Get recall alerts
Free email alert whenever NRT X-RAY A/S has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: NRT X-RAY A/S