Oakworks Inc recalls Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasoun…
- Recall date
- June 29, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2349-2021
- FDA classification
- Class II
- Brand / firm
- Oakworks Inc
- Sold / distributed
- Nationwide
Why it was recalled
Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219