Medical device recalls Moderate risk

Oakworks Inc recalls Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasoun…

Recall date
June 29, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2349-2021
FDA classification
Class II
Brand / firm
Oakworks Inc
Sold / distributed
Nationwide

Why it was recalled

Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219