Medical device recalls Moderate risk

Oculus Optikgeraete GMBH recalls Pentacam AXL REF 70100, Pentacam HR REF 70900. Used to image the anterior segment of the eye.

Recall date
July 8, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1673-2022
FDA classification
Class II
Brand / firm
Oculus Optikgeraete GMBH
Sold / distributed
U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A

Why it was recalled

Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Pentacam AXL REF 70100, Pentacam HR REF 70900. Used to image the anterior segment of the eye.

Get recall alerts

Free email alert whenever Oculus Optikgeraete GMBH has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Oculus Optikgeraete GMBH