Medical device recalls Moderate risk

Giraffe Incubator The Giraffe Incubator is an infant incubator recalled over fire hazard

Recall date
November 18, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Ohmeda Medical recalls Giraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neona…
Recall number
Z-1502-2017
FDA classification
Class II
Brand / firm
Ohmeda Medical
Sold / distributed
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

Why it was recalled

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Giraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed temperature controlled environment