Medical device recalls Moderate risk

Olympus Scientific Solutions Americas recalls Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence

Recall date
July 11, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2760-2017
FDA classification
Class II
Brand / firm
Olympus Scientific Solutions Americas
Sold / distributed
Worldwide Distribution - USA (nationwide) Distribution

Why it was recalled

Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence