OraSure Technologies, Inc. recalls OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola…
- Recall date
- July 20, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2403-2016
- FDA classification
- Class I
- Brand / firm
- OraSure Technologies, Inc.
- Sold / distributed
- Distributed to one customer only in US state of Georgia.
Why it was recalled
Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.