Medical device recalls High risk

OraSure Technologies, Inc. recalls OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola…

Recall date
July 20, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2403-2016
FDA classification
Class I
Brand / firm
OraSure Technologies, Inc.
Sold / distributed
Distributed to one customer only in US state of Georgia.

Why it was recalled

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.