ECMO Reinforced Dual Lumen Cannula recalled over mold contamination
- Recall date
- January 17, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- OriGen Biomedical, Inc. recalls ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV16F
- Recall number
- Z-1227-2019
- FDA classification
- Class I
- Brand / firm
- OriGen Biomedical, Inc.
- Sold / distributed
- US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland…
Why it was recalled
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV16F