OriGen Biomedical, Inc. recalls The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of…
- Recall date
- August 2, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0021-2018
- FDA classification
- Class I
- Brand / firm
- OriGen Biomedical, Inc.
- Sold / distributed
- US Nationwide and the countries of Austria, Belgium, Chile, Czech Republic, Italy, Kingdom of Saudi Arabia, Netherlands, Poland, Sweden, Thailand, Colombia, India, South Africa, United Arab Emirates.
Why it was recalled
Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.
Get recall alerts
Free email alert whenever OriGen Biomedical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: OriGen Biomedical, Inc.