Medical device recalls

Ortho-Clinical Diagnostics recalls The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c R…

Recall date
April 19, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2804-2016
FDA classification
Class III
Brand / firm
Ortho-Clinical Diagnostics
Sold / distributed
Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venez…

Why it was recalled

There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Get recall alerts

Free email alert whenever Ortho-Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ortho-Clinical Diagnostics