Medical device recalls Moderate risk

Ortho-Clinical Diagnostics recalls VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product…

Recall date
November 3, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0640-2016
FDA classification
Class II
Brand / firm
Ortho-Clinical Diagnostics
Sold / distributed
Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, T…

Why it was recalled

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Get recall alerts

Free email alert whenever Ortho-Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ortho-Clinical Diagnostics