Ortho-Clinical Diagnostics recalls VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagn…
- Recall date
- November 19, 2014
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0968-2015
- FDA classification
- Class II
- Brand / firm
- Ortho-Clinical Diagnostics
- Sold / distributed
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Why it was recalled
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).