Ortho-Clinical Diagnostics recalls VITROS 5600 Integrated System- (Refurbished). For use in the in vitro quantitative, semi-quantitative, and qualitative…
- Recall date
- August 15, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0080-2019
- FDA classification
- Class II
- Brand / firm
- Ortho-Clinical Diagnostics
- Sold / distributed
- Nationwide Foreign: France, Brazil, Chile, Great Britain, Phillippines, Spain
Why it was recalled
Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITROS 5600 Integrated System- (Refurbished). For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, Catalog Number: 6802915
Get recall alerts
Free email alert whenever Ortho-Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ortho-Clinical Diagnostics