Ortho-Clinical Diagnostics recalls VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in…
- Recall date
- February 16, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1348-2016
- FDA classification
- Class II
- Brand / firm
- Ortho-Clinical Diagnostics
- Sold / distributed
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Why it was recalled
The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situation where a module that is intended to be de-energized will remain energized only if the associated AC power cord is unplugged.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety of analytes of clinical interest.