Medical device recalls Moderate risk

Ortho Clinical Diagnostics recalls VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Us…

Recall date
July 17, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2771-2020
FDA classification
Class II
Brand / firm
Ortho Clinical Diagnostics
Sold / distributed
Worldwide distribution - US Nationwide distribution.

Why it was recalled

The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.