Medical device recalls Moderate risk

Ortho Clinical Diagnostics recalls VITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage…

Recall date
July 17, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2770-2020
FDA classification
Class II
Brand / firm
Ortho Clinical Diagnostics
Sold / distributed
Worldwide distribution - US Nationwide distribution.

Why it was recalled

The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.

Get recall alerts

Free email alert whenever Ortho Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ortho Clinical Diagnostics