Medical device recalls Moderate risk

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20-…

Recall date
May 8, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2129-2023
FDA classification
Class II
Brand / firm
Ortho-Clinical Diagnostics
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden , The Netherlands, United Kingdom.

Why it was recalled

Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM. Product Code: 6801704

Get recall alerts

Free email alert whenever Ortho-Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.