Medical device recalls Moderate risk

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitr…

Recall date
June 12, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0273-2018
FDA classification
Class II
Brand / firm
Ortho-Clinical Diagnostics
Sold / distributed
Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.

Why it was recalled

The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.