Medical device recalls Moderate risk

ORTHO-CLINICAL DIAGNOSTICS recalls VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Penc…

Recall date
March 7, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1765-2017
FDA classification
Class II
Brand / firm
ORTHO-CLINICAL DIAGNOSTICS
Sold / distributed
Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.

Why it was recalled

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Get recall alerts

Free email alert whenever ORTHO-CLINICAL DIAGNOSTICS has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: ORTHO-CLINICAL DIAGNOSTICS