Ortho Clinical Diagnostics recalls VITROS XT3400 Chemistry System, Product code 6844458
- Recall date
- July 20, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2878-2020
- FDA classification
- Class II
- Brand / firm
- Ortho Clinical Diagnostics
- Sold / distributed
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Why it was recalled
When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITROS XT3400 Chemistry System, Product code 6844458
Get recall alerts
Free email alert whenever Ortho Clinical Diagnostics has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ortho Clinical Diagnostics