Medical device recalls Moderate risk

Ossur Americas recalls OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJ…

Recall date
June 1, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1896-2021
FDA classification
Class II
Brand / firm
Ossur Americas
Sold / distributed
U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Not provided.

Why it was recalled

Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101