Medical device recalls Moderate risk

Ossur H / F recalls LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustab…

Recall date
March 4, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1288-2015
FDA classification
Class II
Brand / firm
Ossur H / F
Sold / distributed
Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,

Why it was recalled

The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30