Medical device recalls Moderate risk

Osteomed, LLC recalls OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Power Control Conso…

Recall date
October 28, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0335-2022
FDA classification
Class II
Brand / firm
Osteomed, LLC
Sold / distributed
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, Guam, HI, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA and WI. The countries of Argentina, Bolivia, Brazil, Canada, Chile, Col…

Why it was recalled

Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Power Control Console 2 REF 450-0021-01 -Power Control Console 2 REF ECOS-450-0021-00 -OsteoMed OsteoPower Power Console 2i REF 450-0005-00 -OsteoMed D-Power Control Console 2i REF 450-0005-01 -Power Control Console 2i REF ECOS-450-0005-00 OsteoMed Footswitches: -OsteoMed BMF Footswitch REF 450-0390 -D-Bi-Directional Master Footswitch REF 450-0380 -D-Uni-Directional Footswitch II REF 450-0350 -D-Bi-Directional Footswitch II REF 450-0360 -D-Uni-Directional Master Footswitch REF 450-0370