Medical device recalls High risk

Pacific Medical Group Inc. recalls Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel, installed by Pacific Medical Group (DBA Avante Health S…

Recall date
March 24, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1375-2021
FDA classification
Class I
Brand / firm
Pacific Medical Group Inc.
Sold / distributed
Distribution US nationwide.

Why it was recalled

Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel, installed by Pacific Medical Group (DBA Avante Health Solutions)

Get recall alerts

Free email alert whenever Pacific Medical Group Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Pacific Medical Group Inc.