Medical device recalls Moderate risk

Partners In Medicine Llc recalls CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic…

Recall date
December 1, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0572-2016
FDA classification
Class II
Brand / firm
Partners In Medicine Llc
Sold / distributed
Nationwide Distribution including MI and CA.

Why it was recalled

Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic parameters time synchronized using Impedance Cardiography (ICG), Electrocardiography (ECG), oscillometric Blood Pressure (oscBP) and continuous Blood Pressure (contBP). Furthermore Sympathic Tone and Parasympathic Tone are calculated by power spectral analysis of ECG (heart rate variability) and continuous blood pressure (blood pressure variability) as well as the Baroreceptor Reflex Sensitivity (BRRS). For use by medically trained personnel. It has been designed for diagnoses aiding and must not be used for vital sign monitoring of critically ill patients under absence of additional, suitable self-monitoring devices or medical trained personnel. The device measures continuously the subject's hemodynamic parameters without reporting any diagnosis. It is the physician's responsibility to make proper judgment based on these parameters.

Get recall alerts

Free email alert whenever Partners In Medicine Llc has a new recall — straight from official government data. Unsubscribe anytime.