Medical device recalls Moderate risk

Permobil recalls PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32P…

Recall date
May 10, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2005-2023
FDA classification
Class II
Brand / firm
Permobil
Sold / distributed
Worldwide distribution.

Why it was recalled

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Get recall alerts

Free email alert whenever Permobil has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Permobil