Phadia US Inc recalls EliA Sample Diluent, Article number, 83-1023-01
- Recall date
- January 30, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1253-2017
- FDA classification
- Class II
- Brand / firm
- Phadia US Inc
- Sold / distributed
- Domestic: AR, CA, CO, FL, GA, IL, IN, KY, LA, MI, MN, MO, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, WI, WV Foreign: None VA/DOD: None
Why it was recalled
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
EliA Sample Diluent, Article number, 83-1023-01