Philips Electronics North America Corporation recalls Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.
- Recall date
- November 3, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0765-2016
- FDA classification
- Class II
- Brand / firm
- Philips Electronics North America Corporation
- Sold / distributed
- Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
Why it was recalled
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.