Philips Electronics North America Corporation recalls CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F syste…
- Recall date
- October 19, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0330-2018
- FDA classification
- Class II
- Brand / firm
- Philips Electronics North America Corporation
- Sold / distributed
- Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
Why it was recalled
The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography