Medical device recalls Moderate risk

Philips Electronics North America Corporation recalls CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F syste…

Recall date
October 19, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0330-2018
FDA classification
Class II
Brand / firm
Philips Electronics North America Corporation
Sold / distributed
Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.

Why it was recalled

The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography