Medical device recalls Moderate risk

Philips Electronics North America Corporation recalls Philips Healthcare Brilliance iCT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to p…

Recall date
January 18, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0019-2018
FDA classification
Class II
Brand / firm
Philips Electronics North America Corporation
Sold / distributed
USA (nationwide) Distribution

Why it was recalled

During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Philips Healthcare Brilliance iCT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Get recall alerts

Free email alert whenever Philips Electronics North America Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Philips Electronics North America Corporation