Medical device recalls Moderate risk

Philips Electronics North America Corporation recalls Philips IntelliVue Module Measurement X2 Model: M3002A

Recall date
January 4, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0854-2016
FDA classification
Class II
Brand / firm
Philips Electronics North America Corporation
Sold / distributed
Worldwide Distribution - Nationwide Distribution and to the countries of : Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Icelan…

Why it was recalled

The ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Philips IntelliVue Module Measurement X2 Model: M3002A