Medical device recalls Moderate risk

Philips Electronics North America Corporation recalls Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453…

Recall date
November 15, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0614-2019
FDA classification
Class II
Brand / firm
Philips Electronics North America Corporation
Sold / distributed
worldwide

Why it was recalled

There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081

Get recall alerts

Free email alert whenever Philips Electronics North America Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Philips Electronics North America Corporation