Philips Medical Systems (Cleveland) Inc recalls BrightView Model 882480 Product Usage: BrightView and BrightView X are gamma cameras designed for single or dual detect…
- Recall date
- December 15, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0512-2018
- FDA classification
- Class II
- Brand / firm
- Philips Medical Systems (Cleveland) Inc
- Sold / distributed
- Worldwide Distribution - US Nationwide
Why it was recalled
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BrightView Model 882480 Product Usage: BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only).
Get recall alerts
Free email alert whenever Philips Medical Systems (Cleveland) Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips Medical Systems (Cleveland) Inc